Heavy Duty moving Surface Probes

Type K Model 213K-TC1-ANP
Type E Model 213E-TC1-ANP
Heavy duty moving surface probes
Temp. Range Tolerance Response time Durability
-50 to 600°C ±2.5°C * 5.0 s B
* at 100°C
Tolerance
■ : Thermocouple type K or E
t : Measured temperature
Model 213■ / 214■ / 215■ / 216■ / 217■
Temperature range Tolerance (Surface)
0°C ≦ t ≦ 200°C ±2.5°C
200°C < t ≦ 300°C ±3.0°C
300°C < t ≦ 400°C ±4.0°C
400°C < t ≦ 500°C ±5.0°C
500°C < t ≦ 600°C ±(0.015╳|t|)°C
The following model designations are used, based on the length of the stainless pipe:
Stainless Pipe Length Model
100mm 213K-TC1-ANP
200mm 214K-TC1-ANP
300mm 215K-TC1-ANP
500mm 216K-TC1-ANP
1000mm 217K-TC1-ANP
Thermocouple type Choose type K or Type E

Be sure to choose the same thermocouple type as the one used for the main thermometer unit.

Cable Length Length of cord section Specify a cord length in 1m increments.
Plug Choose ANP (US Mini Connector) or ASP (Japanese (Anritsu) Standard))

Be sure to choose the same plug type as the one used for the main thermometer unit.

Attention
The following probes have certain restrictions on customization.
Model No. Restrictions
619, 620, 621 These probes are Type K only.