Sick Dusthunter SP100
Probe design with forward scattered light measurement
The SICK Dusthunter SP100 particulate analyzer is just one example of the type of instruments and analyzers that are available to you when you select Cemtek KVB-Enertec as your continuous air monitoring partner and integrator. Cemtek will engineer, source, install and certify any brand, or models needed for your specific application. For more information, see the Download Tab below, or contact us at 504-316-9527, or email [email protected]
[su_tabs active=”4″][su_tab title=”Overview”]
- One-side installation
- For very low to medium dust concentrations
- Automatic check of zero and reference point
- Contamination check
- Hastelloy probe available for corrosive gas environments
- For small to medium duct diameters
- Ideal for thick- or double-walled ducts
- Approved according to EN 15267
- Low maintenance due to self-monitoring and contamination check
- Quick installation – no alignment required
[/su_tab] [su_tab title=”Specifications”]
Description | Scattered light dust measuring devices | ||||||||||||||||
Measured values | Scattered light intensity, dust concentration (after gravimetric comparison measurement) | ||||||||||||||||
Performance-tested measurands | Scattered light intensity | ||||||||||||||||
Measurement principles | Scattered light forward | ||||||||||||||||
Process temperature | Standard version DHSP-T2xx: –40 °C … +220 °C
High temperature version DHSP-T2xx: –40 °C … +400 °C |
||||||||||||||||
Process pressure | With MCU-P control unit: –50 hPa … 10 hPa
With external purge air unit: –50 hPa … 30 hPa With instrument air (provided by the customer): –50 hPa … 100 hPa |
||||||||||||||||
Duct diameter | ≥ 0.25 m | ||||||||||||||||
Conformities | Approved for plants requiring approval
2001/80/EC (13. BImSchV) 2000/76/EC (17. BImSchV) 27.BImSchV TA-Luft (Prevention of Air Pollution) EN 15267 EN 14181 MCERTS 2010/75/EU U.S. EPA PS-11 compliant |
||||||||||||||||
Enclosure rating | IP 66 | ||||||||||||||||
|
|||||||||||||||||
Test functions | Automatic self-test (linearity, contamination, drift, aging)
Contamination limit values: Warning at 30%, fault at 40% Manual linearity test with reference filter |
[/su_tab] [su_tab title=”Applications”]
- Emissions monitoring of power plants and waste incineration plants
- Monitoring of filter systems
- Measurements of exhaust gas or exhaust ventilation ducts
applicationS
[/su_tab] [su_tab title=”Downloads”]Literature and Manual
[/su_tab] [/su_tabs]