| Tape width and shape | 1 | Standard type : Sharp | |
| 2 | Standard type : Flat | ||
| 3 | Narrow type : Sharp | ||
| 4 | Narrow type : Flat | ||
| Coating type and operating temperature range |
1 | Low temperature : Teflon coating (open) | |
| 2 | Low temperature : All teflon coating | ||
| 3 | High temperature : Polyimide coating (open) | ||
| 4 | High temperature : All polyimide coating | ||
| 350°C if the tape length exceeds 150mm. | |||
| Thermocouple | K | Chromel-Alumel | |
| E | Chromel-Constantan | ||
| Tape length | 003 | 30mm | |
| 005 | 50mm | ||
| 008 | 80mm | ||
| 010 | 100mm | ||
| … | …… | ||
| 200 | 2000mm | ||
| Cable type see cable page |
TS | For ST- *1, ST- *2 | |
| GW | For ST- *3, ST- *4 | ||
| 0 | Without cable | ||
| Cable length | 1 | 1000mm | |
| 1.5 | 1500mm | ||
| 2 | 2000mm | ||
| … | …… | ||
| No cable if the cable type is “0.” | |||
| Plug see plug page |
ANP | ||
| ASP | |||
| W | |||
You can choose junction from as below.
Tape width : Standard type
Tape width : Narrow type
You can choose coating type as below.
Low temperature : Teflon coating (Open)
ST-*1 Temp.range : -50 to 210°C
ST-*1 Temp.range : -50 to 210°C

Low temperature : All teflon coating
ST-*2 Temp.range : -50 to 210°C
ST-*2 Temp.range : -50 to 210°C

High temperature : Polyimide coating (Open)
ST-*3 Temp.range : -50 to 400°C
ST-*3 Temp.range : -50 to 400°C
Max. temperature of 350℃ when the tape length exceeds 150mm

High temperature : All polyimide coating
ST-*4 Temp.range : -50 to 400°C
ST-*4 Temp.range : -50 to 400°C
Max. temperature of 350℃ when the tape length exceeds 150mm

You can choose tape length as below.
| 003 : 30mm | 030 : 300mm | 080 : 800mm |
| 005 : 50mm | 035 : 350mm | 090 : 900mm |
| 008 : 80mm | 040 : 400mm | 100 : 1000mm |
| 010 : 100mm | 045 : 450mm | 150 : 1500mm |
| 015 : 150mm | 050 : 500mm | 200 : 2000mm |
| 020 : 200mm | 060 : 600mm | |
| 025 : 250mm | 070 : 700mm |
Specifications
| Tolerance | Open : ±2.5°C (at 100°C) All coating : ±2.5°C (at 100°C) |
| Response time | Open : 2.0 s All coating : 8.0 s |
| Durability rank | S |
Attention
Where the probes are not glued or inserted, the operating temperature limits as follows:
| ST-*1 | ST-*2 | ST-*3 | ST-*4 | |
| Tape ≦ 150mm | 200°C | 180°C | 200°C | 200°C |
| Tape > 150mm | 200°C | 180°C | 180°C | 180°C |
Tolerance
* : Number
■ : Thermocouple type K or E
t : Measured temperature
■ : Thermocouple type K or E
t : Measured temperature
ST-*1■ / ST-*2■
| Temperature range | Tolerance (Internal) |
| -50°C ≦ t < -40°C | ±3.5°C |
| -40°C ≦ t ≦ 210°C | ±2.5°C |
ST-*3■ (Tape length ≦ 150mm)
ST-*4■ (Tape length ≦ 150mm)
ST-*4■ (Tape length ≦ 150mm)
| Temperature range | Tolerance (Internal) |
| -50°C ≦ t < -40°C | ±3.5°C |
| -40°C ≦ t < 333°C | ±2.5°C |
| 333°C ≦ t ≦ 400°C | ±(0.0075╳|t|)°C |
ST-*3■ (Tape length > 150mm)
ST-*4■ (Tape length > 150mm)
ST-*4■ (Tape length > 150mm)
| Temperature range | Tolerance (Internal) |
| -50°C ≦ t < -40°C | ±3.5°C |
| -40°C ≦ t < 333°C | ±2.5°C |
| 333°C ≦ t ≦ 350°C | ±(0.0075╳|t|)°C |